Incident · Recall
Intuitive recalls da Vinci 8mm Monopolar Curved Scissors over extension-tube cracks that can drop fragments into patients
Certain lots of the da Vinci S/Si and X/Xi 8mm Monopolar Curved Scissors were susceptible to cracks near the proximal edge of the extension tube; a crack could fracture the plastic around the instrument shaft and let fragments fall, potentially into the patient. Intuitive issued an Urgent Medical Device Notification advising pre-use inspection.
Distinct from the 2013 Hot Shears tip-insulation recall. Source: FDA Class 2 device recalls Z-3024-2018 (S/Si, 23,064 units) and Z-3025-2018 (X/Xi, 42,274 units), accessdata.fda.gov.
Occurred 2018-08-15 · Intuitive Surgical · Investigated by FDA
How to read this record
Incident records on DEPLOY are compiled from public sources (regulatory filings, news reports, and operator disclosures) and reflect what has been reported and tracked to date. They are not legal findings, determinations of fault, or safety ratings, and may be updated as new information is verified. See the sources below for the underlying references.Machine-readable surfaces
- Markdown mirror: /incidents/davinci-monopolar-scissors-extension-tube-crack-recall-2018.md
- JSON-LD: embedded in this page’s head
- REST API: /v1/incidents/997beef3-1ed9-4990-a1ed-b6e560ea19b8
- Data documentation: /data
- Query this programmatically: Deploy MCP
Sources (1)
Common questions
What happened in Intuitive recalls da Vinci 8mm Monopolar Curved Scissors over extension-tube cracks that can drop fragments into patients?
When did this incident occur?
Who was involved in Intuitive recalls da Vinci 8mm Monopolar Curved Scissors over extension-tube cracks that can drop fragments into patients?
Has anyone responded to Intuitive recalls da Vinci 8mm Monopolar Curved Scissors over extension-tube cracks that can drop fragments into patients?
What is the current status of Intuitive recalls da Vinci 8mm Monopolar Curved Scissors over extension-tube cracks that can drop fragments into patients?
Methodology: Verified · 1 source (1 primary) · last reviewed 2026-08-09
Verification posture
Verified
High confidence
Review state
Stable
Last reviewed 2026-08-09
Sources by quality tier
- 1
- primary-fda-database
- FDA database
The framework is documented at /methodology. Corrections at /corrections. Reviewer: DEPLOY editorial team.
Methodology surface for Intuitive recalls da Vinci 8mm Monopolar Curved Scissors over extension-tube cracks that can drop fragments into patients.Deploy Watch
Track this incident.
We notify you when the regulator updates the record, the remedy status changes, or a related incident surfaces.
Canonical ID 997beef3-1ed9-4990-a1ed-b6e560ea19b8