Vocera Badge auto-acknowledged second critical alarm notification when two arrived within seconds, leaving nurse unaware of second patient's desaturation
Two patients cared for by the same nurse had critical desaturations within seconds of one another. The monitoring system routed alarm notifications for both patients to the same Vocera badge, but the badge did not enunciate the second notification or require a separate acknowledgement. When the nurse acknowledged the first alarm, the second was automatically cancelled.
A passing staff member heard the room alarm and responded, preventing patient harm. Vocera validated the issue. The event was recreated by pulling two code blue buttons simultaneously.
Occurred 2019-04-22 · Vocera Badge · Vocera Communications · Investigated by FDA
How to read this record
Incident records on DEPLOY are compiled from public sources (regulatory filings, news reports, and operator disclosures) and reflect what has been reported and tracked to date. They are not legal findings, determinations of fault, or safety ratings, and may be updated as new information is verified. See the sources below for the underlying references.Machine-readable surfaces
- Markdown mirror: /incidents/vocera-badge-alarm-auto-acknowledge-2019.md
- JSON-LD: embedded in this page’s head
- REST API: /v1/incidents/b52058bb-f0e0-4917-944d-daf2e6a25ae3
- Data documentation: /data
- Query this programmatically: Deploy MCP
Sources (1)
- FDA MAUDE · https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/detail.cfm?mdrfoi__id=8591239&pc=MSX · 2019-05-03
Status history
- Active · 2026-07-30 · agent:recon
Common questions
What happened in Vocera Badge auto-acknowledged second critical alarm notification when two arrived within seconds, leaving nurse unaware of second patient's desaturation?
When did this incident occur?
What robot was involved in Vocera Badge auto-acknowledged second critical alarm notification when two arrived within seconds, leaving nurse unaware of second patient's desaturation?
Has anyone responded to Vocera Badge auto-acknowledged second critical alarm notification when two arrived within seconds, leaving nurse unaware of second patient's desaturation?
What is the current status of Vocera Badge auto-acknowledged second critical alarm notification when two arrived within seconds, leaving nurse unaware of second patient's desaturation?
Methodology: Unreviewed · 1 source (1 primary) · last reviewed 2026-07-30
Verification posture
Unreviewed
Low confidence
Review state
Stable
Last reviewed 2026-07-30
Sources by quality tier
- 1
- primary-fda-database
- FDA database
The framework is documented at /methodology. Corrections at /corrections. Reviewer: DEPLOY editorial team.
Methodology surface for Vocera Badge auto-acknowledged second critical alarm notification when two arrived within seconds, leaving nurse unaware of second patient's desaturation.Deploy Watch
Track this incident.
We notify you when the regulator updates the record, the remedy status changes, or a related incident surfaces.
Canonical ID b52058bb-f0e0-4917-944d-daf2e6a25ae3