CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability

CMR Surgical initiated a Class 2 recall (Z-2717-2026) of two Versius Surgeon Console units (Model V-SS-1020) because Secure Boot was mistakenly not enabled at manufacturing time, presenting a potential cybersecurity risk. The affected units were distributed in California; the firm began notifying customers on June 16, 2026, and will proactively configure Secure Boot on the affected consoles.

Occurred 2026-06-09 · CMR Surgical · Investigated by FDA

How to read this record

Incident records on DEPLOY are compiled from public sources (regulatory filings, news reports, and operator disclosures) and reflect what has been reported and tracked to date. They are not legal findings, determinations of fault, or safety ratings, and may be updated as new information is verified. See the sources below for the underlying references.

Machine-readable surfaces

Sources (1)

  1. FDA · https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=99280 · 2026-07-09

Status history

  • Active · 2026-08-17 · agent:recon

Common questions

What happened in CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?
CMR Surgical initiated a Class 2 recall (Z-2717-2026) of two Versius Surgeon Console units (Model V-SS-1020) because Secure Boot was mistakenly not enabled at manufacturing time, presenting a potential cybersecurity risk. The affected units were distributed in California; the firm began notifying customers on June 16, 2026, and will proactively configure Secure Boot on the affected consoles.
When did this incident occur?
The incident is recorded as occurring on June 9, 2026 on the DEPLOY registry. The date reflects the underlying real-world event, not the registry record's creation date.
Who was involved in CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?
The incident is recorded as involving CMR Surgical on the DEPLOY registry. No specific robot model is linked to this incident in the registry.
Has anyone responded to CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?
No responses to this incident are recorded on the DEPLOY registry. Operators, manufacturers, or affected parties can submit responses to the editorial team; absence is not a guarantee no response was issued.
What is the current status of CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability?
This incident is an active record on the DEPLOY registry; no retraction or correction has been issued.
Methodology: Unreviewed · 1 source (1 primary) · last reviewed 2026-08-17

Verification posture

Unreviewed

Low confidence

Review state

Stable

Last reviewed 2026-08-17

Sources by quality tier

1
primary-fda-database
FDA database

The framework is documented at /methodology. Corrections at /corrections. Reviewer: DEPLOY editorial team.

Methodology surface for CMR Surgical recalls Versius Surgical System consoles due to Secure Boot cybersecurity vulnerability.

Deploy Watch

Track this incident.

We notify you when the regulator updates the record, the remedy status changes, or a related incident surfaces.


Canonical ID d09c1728-a510-4116-86b0-fe0f3a9dd642